A The maintenance tips are categorized day-wise, monthly, bi-annually, and annually. Let us go through each of them in detail:
1 Daily check items 1.1. Check whether the safety induction switch found at the back door of the X-ray machine is working properly.
Method: Close both front and back doors to see whether the contact indicator of the relay is ignited. If the contact indicator light on the relays KA2(front door) and KA3(back door) is on, the safety induction switch is effective. If the indicator light is not on, the safety induction switch is considered as unacceptable.
1.2. Check whether the X-ray machine rotation axis and four-axis movement are working properly.
Method: Once when the total power is turned on, click on the software, and shake the joystick on the comfort to see whether the motion of X-axis and Y-axis is normal. Click both up and down buttons of the X-ray tube and the image detector on the software interface to check if the two axes are moving properly.
1.3. Verify if the motion axis of the X-ray machine sensors is working appropriately.
Method: This test is done to control the movement of each axis to the position of the equivalent limit sensor. When the indicator light of the sensor is switched on, the movement will halt. Check whether the computer is booting properly.
2 Monthly check items 2.1. Inspection of the X-ray machine movement mechanism
● Check whether the movement of each axis is normal and pay attention to the exact method of daily inspection items.
● Check whether the movement of each axis is smooth.
● Check whether the moving parts of the screw nut along with the work platform above the screw are not loose
● Check whether the X-axis and Y-axis of the screw is smooth. If there is any rust formation, apply grease in each of the screws. Check the movement after this lubrication.
2.2. Inspection of the emergency stop switch working condition
● Press the emergency stop button manually to check if the X-ray machine is down.
3 Biannual check items 3.1. X-ray machine power detection
Method: First, open the back door. Check whether 24VDC and 5VDC power supplies are working properly.
3.2. Equipment guide rail on the lubricating oil for its clean appearance
Method: First, turn off the device as well as disconnect the power supply. Open the back door, lubricate each screw rod with grease, and run it over and over to lubricate while cleaning the portions of the inner wall. In the end, open the side door and wipe it with a clean cloth.
4 Annual maintenance 4.1. Check for the signs of damage on the inner side and the outer side of the machine, if any.
4.2. Check the observation window glass to find if there is any scratch.
4.3. Check the rotation pattern of the X-ray light tube, platform movement axis, as well as the detector.
4.4. Check all the fans and clean the filter properly with cotton.
5 Check X-ray image quality Description of the problem | Analysis of the problem | Solutions |
The computer has a starting problem | · Loose computer power connectors · Presence of virus can cause a system failure | · Reinsert your computer’s power plug · Reinstall the entire system |
Automatic reboot after power-up | Presence of the virus in the computer | Perform system restoration |
X/Y/Z axis motors are not running properly | · Issues with the power supply to the drive · Loose B joints · Drive anomalies | · Replace the power supply · Use fasteners · Use replacement Drive |
Jarring of the rail during the movement | · Loose couplings · Linear guide and roller screw without any lubrication for a longer duration due to dryness | · Check couplings and tighten them. · Lubricate the linear guides and roller screws using grease |
Alarmed UPS | Inadequate power supply | Change the power supply |